DRX-Revolution Mobile X-Ray System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
DRX-Revolution Mobile System |
More than 10 numbers, contact manufacturer. |
8618894 |
DRX-Revolution Mobile X-Ray System |
More than 10 numbers, contact manufacturer. |
8610180 |
Issue
There is a potential for unexpected failure of the electrical components within the Carestream 1.0 generator installed in your DRX-Revolution Mobile X-Ray System(s). This can lead to temporary and self-contained thermal overload within the generator. Should an event like this occur, the system will become inoperable and loud noise, burnt smell and smoke may be detectable.
Recall start date: September 20, 2023
Additional information
Details
Carestream Health, Inc.
150 Verona Street, Rochester, New York, United States, 14608
Get notified