DRX-Revolution Mobile X-Ray System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
DRX-Revolution Mobile X-Ray System |
More than 10 numbers, contact manufacturer. |
1019397 |
Issue
There is a potential for unexpected failure of the electrical components within the Communication & Power Industries, Inc (CPI) generator installed in the DRX-revolution mobile X-ray system(s). This can lead to temporary and self-contained thermal overload within the generator. Should an event like this occur, the system will become inoperable and loud noise, burnt smell and smoke may be detectable.
Recall start date: September 20, 2023
Additional information
Details
Carestream Health, Inc.
150 Verona Street, Rochester, New York, United States, 14608
Get notified
Receive notifications for new and updated recalls and alerts by category.