Health product recall

Drill Sleeve Guide (2018-05-02)

Starting date:
May 2, 2018
Posting date:
May 25, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66852

Affected products

Drill Sleeve Guide

Reason

The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.

Affected products

Drill Sleeve Guide

Lot or serial number

105176 613121
107104 710220
111423 716180
113775 719190
113776 719190
114390 715001
114724 711101
115000 714201
115847 712211
116389 712121
117146 810110

Model or catalog number

DSG-6.3-090-2.4N

Companies
Manufacturer

Ad-Tech Medical Instrument Corp.

1901 William St.

Racine

53404

Wisconsin

UNITED STATES