DR 800
Brand(s)
Last updated
Summary
Product
DR 800
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
DR 800 | Not applicable. | 6010/200 |
Issue
A customer reported a broken front chain on the DR 800 and the spring package came out from the table. There was no reported harm to users or patients during this event. There have been no additional reported events. The supplier determined the front lever chain of the table might present a reduction in reliability over time. The potential risks have been evaluated by the supplier and the supplier initiated a field safety corrective action.
Recall Start Date: July 29, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Agfa N.V.
Septestraat 27, Mortsel, Belgium
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75931
Get notified