This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DOVER MIDSTREAM SPECIMEN CATCH KIT (2016-01-25)

Starting date:
January 25, 2016
Posting date:
February 12, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57108

Affected Products

A. DOVER MIDSTREAM SPECIMEN CATCH KIT

Reason

Covidien received notification from third party manufacturer Aplicare of a quality issue related to specific production lots of Aplicare's Castile Soap Towelettes for which Aplicare has initiated a recall. Aplicare has received user complaints stating a perceivable malodor from the soap which has been determined to be related to the presence of Nesterenkonia lacusekhoensis in the soap.

Affected products

A. DOVER MIDSTREAM SPECIMEN CATCH KIT

Lot or serial number

Not applicable

Model or catalog number
  • 2090SA
  • 4090SA
Companies
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES