Health product recall

DLP Cardiac Vents & Sumps

Brand(s)
Last updated

Summary

Product
DLP Cardiac Vents & Sumps
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

DLP Cardiac Vents & Sumps

More than 10 numbers, contact manufacturer.

12115

DLP Cardiac Vents & Sumps

More than 10 numbers, contact manufacturer.

12110

DLP Cardiac Vents & Sumps

More than 10 numbers, contact manufacturer.

12113

Issue

Since March 2024, there has been an increase in complaints reported for the DLP malleable left heart vent catheter for the catheters resisting shape retention when being bent. The catheters are intended to be malleable and retain a bend in the shaft. It was determined that the primary supplier updated their specification to remove annealing process to attain tensile strength listed in ASTM1313-18.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Medtronic Inc.

710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78123

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe