This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Distalizer Jet Appliance Lock Nut (2014-01-03)

Starting date:
January 3, 2014
Posting date:
February 11, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37919

Recalled products

  1. Distalizer Jet Appliance Lock Nut 

Reason

A lock nut used to tighten the wire within the Distal jet device was missing the screw collar, which serves as a mechanical stop to prevent over tightening.

Affected products

A. Distalizer Jet Appliance Lock Nut 

Lot or serial number
  • 2013.08.13 - 2013.11.06
Model or catalog number
  • 3000-0098
Companies
Manufacturer

Allesee Orthodontic Appliances, Inc.

13931 Spring Street

Sturtevant

53177

Wisconsin