Product
Disposable Medical Isolation Gowns Level 2
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Disposable Medical Isolation Gowns Level 2 |
>10 Numbers, contact mfg |
T0202 |
Issue
Samuelsohn Medical Limited has been informed that certain lots of level 2 disposable medical isolation gowns have failed the quality control test for hydrostatic pressure along the seams (aatc tm127-2017).
Recall Start Date: June 22, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Samuealsohn Medical Limited.
6930 Avenue Du Parc, Montreal, Quebec, Canada, H3N 1W9
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73916