This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Disposable Fiber Optic Medium Laryngoscope Handle

Starting date:
July 14, 2017
Posting date:
August 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64138

Reason

Due to handles requiring batteries to operate, it was determined the potential for insufficient contact causing inconsistent (intermittent or flickering) lighting was present for a small number of the handles. Intermittent lighting may compromise the user's ability to visualise the patient's airway which may result in delayed treatment.

Affected products

Disposable Fiber Optic Medium Laryngoscope Handle

Lot or serial number

  • 2216
  • 3016
  • 4016
  • 4816

Model or catalog number

OL-334L6

Companies

Manufacturer

Bomimed Inc.

96 Terracon Place

Winnipeg

Manitoba

CANADA