Discovery, Signa and Optima MR Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Discovery Mr750 3.0t |
902473MR2 |
DISCOVERY MR750 3.0T |
Discovery Mr450 1.5t Mr System |
250727MR2 |
DISCOVERY MR450 1.5T MR SYSTEM |
Discovery Mr750w 3.0t |
613737MR1 |
DISCOVERY MR750W 3.0T |
Signa Pet/Mr |
604822MRP1 |
SIGNA PET/MR |
Signa Pioneer |
403253MR1 |
SIGNA PIONEER |
Signa Architect |
905527MR1 |
SIGNA ARCHITECT |
Signa Hdxt 1.5t Mr System |
514473MR1 |
SIGNA HDXT 1.5T MR SYSTEM |
Signa Hdxt 3.0t Mr System - Main Unit |
416369MR3T |
SIGNA HDXT 3.0T MR SYSTEM |
Optima Mr450w Mri System - Main Unit - 16ch |
902893MR1 |
OPTIMA MR450W 1.5T |
Optima Mr450w Mri System - Main Unit - 32ch |
613938MR1 |
M7004ZJ |
Optima Mr450w Magnetic Resonance Imaging System |
More than 10 numbers, contact manufacturer. |
M7000YN |
Signa Premier |
514934MR4 |
SIGNA PREMIER |
Issue
B1+RMS predicted value may exceed user prescribed limit in low SAR mode for certain MR systems.
Recall start date: June 14, 2024
Additional information
Details
Ge Medical Systems Llc |
3200 North Grandview Blvd, Waukesha, Wisconsin, United States, 53188 |