Discovery Nm/Ct (2017-06-01)
- Starting date:
- June 1, 2017
- Posting date:
- June 13, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-63600
Affected Products
- A. Discovery Nm/Ct 570c
- B. Discovery Nm/Ct 670 Pro
- C. Discovery Nm/Ct 670
- D. Discovery Nm/Ct 670 Es
Reason
GE Healthcare has recently become aware of a potential safety issue that can result in a scan starting before the prescribed delay has completed. More specifically, if an exam is interrupted during the prep delay period using the emergency stop button and the scan is then resumed, the delay period may be shorter than expected. If personnel are in the scan room at the time of the premature scan initiation they can be exposed to scatter x-ray radiation.
Affected products
A. Discovery Nm/Ct 570c
Lot or serial number
00000000019102
Model or catalog number
S8000RD
Companies
- Manufacturer
-
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL
B. Discovery Nm/Ct 670 Pro
Lot or serial number
000000SAN00776
27051
27081
27115
27124
PROV27146
Model or catalog number
H3100AD
Companies
- Manufacturer
-
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL
C. Discovery Nm/Ct 670
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
H3100JN
Companies
- Manufacturer
-
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL
D. Discovery Nm/Ct 670 Es
Lot or serial number
28008
Model or catalog number
H3100AE
Companies
- Manufacturer
-
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL