Health product recall

Discovery MR450 Narrow Bore 1.5T MRI

Starting date:
January 26, 2017
Posting date:
February 9, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62136

Reason

Potential safety issue with the patient bore heating on the Discovery MR450 Narrow Bore 1.5T MRI. In the event of radio frequence (RF) component failure, typically happening after extended system use, a small area on the bore surface may become warmer than normal. If this happens in an area where there is patient contact with the bore, and proper padding is not used, it may cause a serious burn.

Affected products

Discovery MR450 Narrow Bore 1.5T MRI

Lot or serial number

  • 000000000R7149
  • 000000000R7320
  • 00000944BD616B
  • 00001101XF5028

Model or catalog number

  • DISCOVERY MR450 1.5T MR SYSTEM

Companies

Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd

Waukesha

53188

Wisconsin

UNITED STATES