This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DISCOVERY IGS 740 - MAIN UNIT (2015-06-02)

Starting date:
June 2, 2015
Posting date:
June 26, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Identification number:
RA-53938

Recalled Product

A.DISCOVERY IGS 740 - MAIN UNIT

Reason

The affected systems may experience a loss of gantry motion capability as a result of a failure with the Gantry Motion Control Board (MCB). Limited gantry motion can interfere, complicate or prolong an interventional procedure dependent upon multiple fields of view. Diagnostic and image guided performance capability may be hindered. After a reset, X- rays are available and manual motions of some gantry and table axes are still available. There have been no injuries reported as a result of this issue.

Affected products

A.DISCOVERY IGS 740 - MAIN UNIT

Lot or serial number

15A2A0007

Model or catalog number

5418813

Companies
Manufacturer

GE Medical Systems SCS

283 Rue De La Miniere

Buc