Diode Laser WOLF 445 nm (2019-03-27)
- Starting date:
- March 27, 2019
- Posting date:
- April 12, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69568
Last updated:
Affected Products
Diode Laser WOLF 445 nm
Reason
A software bug was detected by the manufacturer's engineering; it may cause higher power output than the one shown on the unit's screen. It is rare and can only happen if the user deliberately uses a shortcut to adjust the power (instead of using the "+" and "-" icons) along with a specific combination of 4 unique laser parameters (which is a again a rare combination). The laser has been in use worldwide for 2 years and no one to date has encountered this.
Affected products
Diode Laser WOLF 445 nm
Lot or serial number
650099
Model or catalog number
LS11000-445
Companies
- Manufacturer
-
A.R.C. laser GmbH
Bessemerstraße 14
Nürnberg
90411
GERMANY