Health product recall

Dimension® Creatinine (CRE2)

Last updated

Summary

Product
Dimension® Creatinine (CRE2)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

Dimension® Creatinine (CRE2)

GA6307

10872079

Dimension® Creatinine (CRE2)

BA7005

10872079

Issue

Siemens Healthineers has confirmed, through investigation of customer complaints, a potential for imprecision in Dimension® Creatinine (CRE2) quality control (QC) and patient sample results when using lot numbers GA6307 and BA7005 on the dimension system.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Siemens Healthcare Diagnostics Inc.

500 Gbc Drive, Newark, Delaware, United States, 19714

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-81833

Get notified

Receive emails about new and updated recall and safety alerts.