This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Dimension Vista System - Enzyme Calibrator (2014-01-20)

Starting date:
January 20, 2014
Posting date:
February 11, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37925

Recalled products

  1. Dimension Vista System - Enzyme Calibrator 

Reason

There have been reports of quality control (QC) shifts with pseudocholinesterase (PCHE) and imprecision during PCHE calibration with Dimension Vista enzyme 1 calibrator (ENZ 1 cal) lots 3FD034 and 3HD012. Investigation by Siemens confirms a loss of PCHE activity in the affected lots of enzyme 1 calibrator, causing an elevation of qc by as much as 80% and in some instances, imprecision of replicates during calibration.

Affected products

A. Dimension Vista System - Enzyme Calibrator 

Lot or serial number
  • 3FD034
  • 3HD012
Model or catalog number
  • KC310
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES