Dimension Vista 500 and Dimension Vista 1500 System Analyzers (2014-04-09)
- Starting date:
- April 9, 2014
- Posting date:
- April 30, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39247
Recalled products
A. Dimension Vista System - Dimension Vista 1500 Analyzer
B. Dimension Vista System - Dimension Vista 500 Analyzer
Reason
A small number of Vista air compressors from a vendor were miswired. If installed and powered up, these components could cause a fire, or become an electrical shock hazard.
Affected products
A. Dimension Vista System - Dimension Vista 1500 Analyzer
Lot or serial number
All lots
Model or catalog number
1000033793
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
B. Dimension Vista System - Dimension Vista 500 Analyzer
Lot or serial number
All lots
Model or catalog number
1000034648
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES