Dimension Sample Probe Cleaner (2018-09-27)
- Starting date:
- September 27, 2018
- Posting date:
- October 19, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67952
Affected products
Dimension Sample Probe Cleaner
Reason
Siemens is notifying customers of the observation of the error message "sample probe cleaner not detected" (error code 569) during system check and leaking of sample probe cleaner bottle caps. Siemens has confirmed a low frequency (less than 1%) of defective bottles resulting in this error.
Affected products
Dimension Sample Probe Cleaner
Lot or serial number
56420006 AND HIGHER
Model or catalog number
RD703
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES