DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D 100 TYPE 5410 OR DX-D 100 WIRELESS TYPE 5411 (2015-06-05)
- Starting date:
- June 5, 2015
- Posting date:
- August 10, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54522
Affected Products
A. DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D 100 TYPE 5410 OR DX-D 100 WIRELESS TYPE 5411
Reason
When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur. Capacitive touch screens rely on the electrical properties of the human body to enable you to operate the device with very light touches of a finger. However, when liquid comes in contact with the touch screen it may incorrectly recognize this as user input.
Affected products
A. DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D 100 TYPE 5410 OR DX-D 100 WIRELESS TYPE 5411
Lot or serial number
> 10 numbers, contact mfg.
Model or catalog number
- SM-40HF-B-D-C
- SM-40HF-B-D-T
- SM-40HF-B-D-V
- SM-50HF-B-D-C
- SM-50HF-B-D-T
- SM-50HF-B-D-V
Companies
- Manufacturer
-
AGFA Healthcare N.V.
Septestraat 27
Mortsel
2640
BELGIUM