Digital Diagnostic Mobile X-Ray System GM85
Brand(s)
Summary
Customers to be contacted to arrange visit for inspection and repair.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Digital Diagnostic Mobile X-Ray System GM85 |
S/N 5143M3UJ200001N |
GM85 |
Issue
We have received a report regarding the loosening or breakage of bolts that connect the tube head unit (THU) to the arm. An investigation determined that this may be caused by metal fatigue resulting from accumulated stress over a long period of use. If all connecting bolts were to fail, it could potentially cause the THU to detach from the arm, posing a risk of serious injury to a patient or user.
Additional information
Details
Samsung Electronics Co., Ltd.
129, Samsung-Ro, Yeongtong-Gu, Suwon-Si, Gyeonggi-Do, South Korea, 16677
Get notified