Digital Diagnost (2018-06-12)
- Starting date:
- June 12, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67406
Affected products
Digital Diagnost
Reason
The footplate of the Patient support for stitching can be fixed in upper position. When the fixing clamp (hook) is not in its correct looking position, the footplate is not securely held. If it falls down, it may hurt the operator's or patient's foot or leg (e.g. when moving the stand or when the operator fails to fix it correctly upon moving the footplate up).
Affected products
Digital Diagnost
Lot or serial number
- More than 10 numbers, contact manufacturer.
Model or catalog number
- 9890 010 83904
- 9890 010 8390X
Companies
- Manufacturer
-
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY