Health product recall

Digital Diagnost (2018-06-12)

Starting date:
June 12, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67406

Affected products

Digital Diagnost

Reason

The footplate of the Patient support for stitching can be fixed in upper position. When the fixing clamp (hook) is not in its correct looking position, the footplate is not securely held. If it falls down, it may hurt the operator's or patient's foot or leg (e.g. when moving the stand or when the operator fails to fix it correctly upon moving the footplate up).

Affected products

Digital Diagnost

Lot or serial number
  • More than 10 numbers, contact manufacturer.
Model or catalog number
  • 9890 010 83904
  • 9890 010 8390X
Companies
Manufacturer
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY