This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Digital Diagnost (2015-01-16)
- Starting date:
- January 16, 2015
- Posting date:
- January 28, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-43479
Recalled products
- Digital Diagnost
Reason
The system is designed to emit an acoustic signal (beep) upon termination of an exposure. However, if the system has been powered on for more than 54 hours, the system will no longer emit this signal.
Affected products
A. Digital Diagnost
Lot or serial number
- SN14000054
Model or catalog number
- 9890 010 83904
Companies
- Manufacturer
-
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY