Digital Diagnost (2015-01-16)

Starting date:
January 16, 2015
Posting date:
January 28, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43479

Recalled products

  1. Digital Diagnost

Reason

The system is designed to emit an acoustic signal (beep) upon termination of an exposure. However, if the system has been powered on for more than 54 hours, the system will no longer emit this signal.

Affected products

A. Digital Diagnost

Lot or serial number
  • SN14000054
Model or catalog number
  • 9890 010 83904
Companies
Manufacturer

Philips Medical Systems DMC GmbH

Rontgenstrabe 24

Hamburg

22335