This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Digital Diagnost (2014-03-20)

Starting date:
March 20, 2014
Posting date:
April 14, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39029

Recalled Products

Digital Diagnost

Reason

The footboard of patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held. If it falls down, it may hurt the operator's or patient's foot (for example - when moving the stand or when the operator fails to fix it correctly upon moving the board up).

Affected products

Digital Diagnost 

Lot or serial number

More than 10 numbers, contact manufacturer 

Model or catalog number

9890 010 8390X

Companies
Manufacturer
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY