Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58576

Affected products 

A. Dialog+ Dialysis Machines

Reason

Flanges assembled into conductivity cells show hairline cracks.  If a crack occurs it may lead to a leakage and result in a balance deviation.  Affected machines will have improper filtration rate.

Affected products

A. Dialog+ Dialysis Machines

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

710200K
710200L
710500L

Companies
Manufacturer
B. Braun Avitum Ag
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY