This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Dialog+ Dialysis Machines (2016-03-18)

Starting date:
March 18, 2016
Posting date:
May 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58576

Affected products 

A. Dialog+ Dialysis Machines

Reason

Flanges assembled into conductivity cells show hairline cracks.  If a crack occurs it may lead to a leakage and result in a balance deviation.  Affected machines will have improper filtration rate.

Affected products

A. Dialog+ Dialysis Machines

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

710200K

710200L

710500L

Companies
Manufacturer

B. Braun Avitum Ag

Schwarzenberger Weg 73-79

Melsungen

34212

GERMANY