This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DIAGNOSTIX 703 PALM ANEROID SPHYGMOMANOMETER (2016-05-16)

Starting date:
May 16, 2016
Posting date:
October 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64768

Affected products

DIAGNOSTIX 703 PALM ANEROID SPHYGMOMANOMETER

Reason

The sphygmomanometer uses a small connector to connect its single tube to the aneroid gauge. This connector was inverted when it was placed on the tube and cannot connect to the gauge in its current orientation.

Affected products

DIAGNOSTIX 703 PALM ANEROID SPHYGMOMANOMETER

Lot or serial number
  • 60419283
  • 60405283
Model or catalog number

703-11ABK

Companies
Manufacturer

AMERICAN DIAGNOSTIC CORPORATION

55 COMMERCE DRIVE

HAUPPAUGE

11788

New York

UNITED STATES