This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
DGPHP RFA High-Power Single Use Grounding Pads and Cool-Tip RFS Electrode Kits
- Starting date:
- August 7, 2012
- Posting date:
- September 3, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Product Safety, Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-15532
Recalled Products
A. DGPHP RFA High-Power Single Use Grounding Pads and Cool-Tip RFS Electrode Kits
Reason
This recall is being conducted due to the potential for foil degradation on the DGPHP RFA High-Power Single Use Grounding Pads. This foil degradation may result in an electrical disconnect and undesirable thermal profile during use which may result in a serious burn at the patient pad site.
Affected products
A. DGPHP RFA High-Power Single Use Grounding Pads and Cool-Tip RFS Electrode Kits
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
More than 10 numbers, contact manufacturer.
Companies
- Manufacturer
-
Covidien LLC