Health product recall

Dermatome An Handpiece (2018-06-12)

Starting date:
June 12, 2018
Posting date:
June 22, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67102

Affected products

Dermatome An Handpiece

Reason

Zimmer Biomet is conducting a medical device recall for specific serial numbers of Dermatome An. The affected devices could potentially have a loose control bar, which could compromise the ability to control the thickness of the graft. The affected devices could also potentially have a loose width plate that if undetected, could result in an imperfect, yet still usable graft.

Affected products

Dermatome An Handpiece

Lot or serial number
  • 700198
  • 700201
  • 700202
  • 700205
  • 700206
  • 700207
  • 700208
  • 700209
Model or catalog number

88710100