Health product recall

Delfia Inducer

Brand(s)
Last updated

Summary

Product
Delfia Inducer
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Delfia Inducer

742762
742802

3027-0010

Issue

On Delfia Inducer bottles (2 kit lots), the label is placed too low preventing the Delfia Xpress instrument level sensor to detect the Delfia inducer liquid level. Because of that the instrument does not give an 'inducer bottle soon empty' warning at level where there is 4 ml Delfia inducer left (30 tests + 1 ml dead volume). This may cause the Delfia inducer to run out and this in turn may cause failed or false results.

Recall start date: February 7, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Wallac Oy

Mustionkatu 6, Turku, Finland, 20750

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72686

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe