Defigard Touch7 Monitor And Defibrillator
Brand(s)
Last updated
Summary
Product
Defigard Touch7 Monitor And Defibrillator
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Defigard Touch7 Monitor And Defibrillator | All lots. | 1-128-9900 |
Issue
In rare cases, an ECG failure affecting impedance measurement via defibrillation electrodes has led to the Defigard Touch-7 falsely reports an electrode fault, resulting in inoperative defibrillation.
Recall start date: November 14, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Schiller Medical |
4 Rue Louis Pasteur, Wissembourg, France, 67160 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-74632
Get notified
Receive notifications for new and updated recalls and alerts by category.