Health product recall

Defigard Touch7 Monitor And Defibrillator

Brand(s)
Last updated

Summary

Product
Defigard Touch7 Monitor And Defibrillator
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Defigard Touch7 Monitor And Defibrillator

All lots.

1-128-9900

Issue

In rare cases, an ECG failure affecting impedance measurement via defibrillation electrodes has led to the Defigard Touch-7 falsely reports an electrode fault, resulting in inoperative defibrillation.

Recall start date: November 14, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Schiller Medical
4 Rue Louis Pasteur, Wissembourg, France, 67160
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-74632

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe