DEFIGARD Touch 7
Brand(s)
Last updated
Summary
Product
DEFIGARD Touch 7
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
DEFIGARD Touch 7 | All lots. | 1-128-9900 |
Issue
In the past three years, six failures in the defibrillation shock delivery control circuit of the DEFIGARD Touch-7 have led to shocks not being delivered when the device is used.
Recall Start Date: June 28, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Schiller Medical
4 Rue Louis Pasteur, Wissembourg, France, 67160
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75829
Get notified
Receive notifications for new and updated recalls and alerts by category.