Health product recall

DEFIGARD Touch 7

Brand(s)
Last updated

Summary

Product
DEFIGARD Touch 7
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

DEFIGARD Touch 7

All lots.

1-128-9900

Issue

In the past three years, six failures in the defibrillation shock delivery control circuit of the DEFIGARD Touch-7 have led to shocks not being delivered when the device is used.

Recall Start Date: June 28, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Schiller Medical

4 Rue Louis Pasteur, Wissembourg, France, 67160

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75829

Get notified

Receive notifications for new and updated recalls and alerts by category.