Health product recall

Decompression Of Choice Table

Last updated

Summary

Product
Decompression Of Choice Table
Issue
Medical devices - Performance
What to do

User will be contacted by manufacturer with instructions to follow. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Decompression Of Choice Table

WTN-DOC04262458487
WTN-DOC04262458489
WTN-DOC04262458488

E9011

Issue

There is possible defectiveness on part E7130 on the decompression table.  There is a lot of 30 pieces of E7130 where rework was done on some of the pieces and the rework led to a weld that may have insufficient penetration.

Recall start date: June 25, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Pivotal Health Solutions

3003 9th Avenue Sw, Watertown, South Dakota, United States, 57201

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75863

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe