Darco System 7.0 Headless Screw (Sterile)
Brand(s)
Last updated
Summary
Product
Darco System 7.0 Headless Screw (Sterile)
Issue
Medical devices - Device compatibility
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Darco System 7.0 Headless Screw (Sterile) | 1733433 | DCS01660 |
Issue
Subsequent to a product complaint (April 8 2024), Stryker confirmed that one specific lot of Darco™ 7.0mm Headless Screws was manufactured without the hex feature required for mating with the associated screwdriver.
Recall start date: September 3, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
Wright Medical Technology Inc. |
1023 Cherry Road, Memphis, Tennessee, United States, 38117 |
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-76033
Get notified