Dako Coverstainers
Last updated
Summary
Product
Dako Coverstainers
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Coverstainer, Dako Coverstainer Cpo |
CS10030,CS10040,CS10026 |
CS100 |
Dako Coverstainer, Dako Coverstainer Cpo |
CS10030,CS10040,CS10026 |
CS100 |
Issue
Lower front door of coverstainer instrument damaged due to hinges not being able to support weight of the door assembly. This leads to cracks in the plexiglass of the lower front door and causes pieces to break off in the top left/right corners of the assembly.
Recall start date: April 2, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
Agilent Technologies Singapore (International) Pte Ltd. |
No. 1 Yishun Avenue 7, Singapore, Singapore, 768923 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75383
Get notified