Health product recall

Dako Coverstainers

Last updated

Summary

Product
Dako Coverstainers
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Coverstainer, Dako Coverstainer Cpo

CS10030,CS10040,CS10026

CS100

Dako Coverstainer, Dako Coverstainer Cpo

CS10030,CS10040,CS10026

CS100

Issue

Lower front door of coverstainer instrument damaged due to hinges not being able to support weight of the door assembly. This leads to cracks in the plexiglass of the lower front door and causes pieces to break off in the top left/right corners of the assembly.

Recall start date: April 2, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
Agilent Technologies Singapore (International) Pte Ltd.
No. 1 Yishun Avenue 7, Singapore, Singapore, 768923
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75383

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe