Dako Autostainer Link 48 with DakoLink (2014-07-01)
- Starting date:
- July 1, 2014
- Posting date:
- August 1, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40797
Recalled Products
Dako Autostainer Link 48 with DakoLink
Reason
Software defect in the calibration routine of DakoLink 4.0, which may cause the auto stainer control software to use old calibration settings, where an instrument has, a preventative maintenance or waste pump replaced since installation of the DakoLink 4.0 software.
Affected products
Dako Autostainer Link 48 with DakoLink
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
AS480
Companies
- Manufacturer
-
Dako North America, Inc.
6392 via Real
Carpinteria
93013
California
UNITED STATES