This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DA VINCI SI (IS3000) SURGEON CONSOLE TOUCHPAD (2016-02-16)

Starting date:
February 16, 2016
Posting date:
March 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57406

Affected Products

DA VINCI SI (IS3000) SURGEON CONSOLE TOUCHPAD

Reason

Intuitive Surgical has found that compact flash cards used in touchpads on specific da Vinci Si and Xi surgical systems have the potential to become corrupted when the touchpads experience an abrupt power loss and/or are powered on and off within a short period of time.

Affected products

DA VINCI SI (IS3000) SURGEON CONSOLE TOUCHPAD

Lot or serial number
  • SH0948
  • SH1104
  • SH2240
Model or catalog number
  • SS3000
Companies
Manufacturer

Intuitive Surgical

1266 Kifer Road, Building 101,

Sunnyvale

94086

California