DA VINCI SI (IS3000) SURGEON CONSOLE TOUCHPAD (2016-02-16)
- Starting date:
- February 16, 2016
- Posting date:
- March 4, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57406
Affected Products
DA VINCI SI (IS3000) SURGEON CONSOLE TOUCHPAD
Reason
Intuitive Surgical has found that compact flash cards used in touchpads on specific da Vinci Si and Xi surgical systems have the potential to become corrupted when the touchpads experience an abrupt power loss and/or are powered on and off within a short period of time.
Affected products
DA VINCI SI (IS3000) SURGEON CONSOLE TOUCHPAD
Lot or serial number
- SH0948
- SH1104
- SH2240
Model or catalog number
- SS3000
Companies
- Manufacturer
-
Intuitive Surgical
1266 Kifer Road, Building 101,
Sunnyvale
94086
California
UNITED STATES