This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Da Vinci S, Si, and Si-e - battery box of patient side cart (2014-08-12)

Starting date:
August 12, 2014
Posting date:
August 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41219

Recalled products

  1. Da Vinci S, Si, and Si-e - battery box of patient side cart

Reason

Patient side cart battery boxes returned from the field revealed that in some rare charging conditions the battery can overheat. Overheating has the potential to result in the battery bulging and the release of internal battery gases into surrounding air.

Affected products

A. Da Vinci S, Si, and Si-e - battery box of patient side cart

Lot or serial number
  • SG359
  • SH0260
  • SH0370
  • SH1104
  • SH1163
  • SH1184
  • SH1319
Model or catalog number
  • 371268-02
Companies
Manufacturer

Intuitive Surgical

1266 Kifer Road, Building 101

Sunnyvale

94086

California

UNITED STATES