Health product recall

Cytomics FC 500 Flow Cytometer (2013-08-28)

Starting date:
August 28, 2013
Posting date:
October 16, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36183

Recalled Products

Cytomics FC 500 Flow Cytometer

Reason

Beckman Coulter has determined that when the maintenance log for December is printed or saved,entries for the subsequent eleven (11) months are lost and cannot be retrieved unless the database was backed-up prior to performing the print or save actions. For example, if December 2012 is printed or saved, any maintenance log entries for January - November 2013 will be lost. This issue can result in potentially incomplete maintenance log entries. No other report generator functions or patient results are affected by this issue.

Affected products

Cytomics FC 500 Flow Cytometer 

Lot or serial number

All lots 

Model or catalog number
  • 626553
  • 6605628
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES