Cytomics FC 500 Flow Cytometer (2013-08-28)
- Starting date:
- August 28, 2013
- Posting date:
- October 16, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36183
Recalled Products
Cytomics FC 500 Flow Cytometer
Reason
Beckman Coulter has determined that when the maintenance log for December is printed or saved,entries for the subsequent eleven (11) months are lost and cannot be retrieved unless the database was backed-up prior to performing the print or save actions. For example, if December 2012 is printed or saved, any maintenance log entries for January - November 2013 will be lost. This issue can result in potentially incomplete maintenance log entries. No other report generator functions or patient results are affected by this issue.
Affected products
Cytomics FC 500 Flow Cytometer
Lot or serial number
All lots
Model or catalog number
- 626553
- 6605628
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES