CyberKnife Treatment Delivery System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
CyberKnife Treatment Delivery System |
C0417 |
0660000 |
Issue
Accuray Incorporated has discovered an issue related to the CyberKnife Standard Treatment Couch (STC). When an internal mechanism becomes disconnected during clinical use, uncontrolled rotation can occur beyond the specified degrees of motion. In rare circumstances snap rings have become partially or completely detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the snap rings are removed or damaged.
Recall Start Date: November 15, 2024