CyberKnife Treatment Delivery System
Brand(s)
Last updated
Summary
Product
CyberKnife Treatment Delivery System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| CyberKnife Treatment Delivery System | C0417 C0545 C0573 C0351 | 0660000 |
Issue
Accuray Incorporated has discovered an issue related to the CyberKnife Standard Treatment Couch (STC). When an internal mechanism becomes disconnected during clinical use, uncontrolled rotation can occur beyond the specified degrees of motion. In rare circumstances snap rings have become partially or completely detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the snap rings are removed or damaged.
Recall Start Date: November 15, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Accuray, Inc.
1209 Deming Way, Madison, Wisconsin, United States, 53717
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76568
Get notified