This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CXI SUPPORT CATHETER (2015­-10-­22)

Starting date:
October 22, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55858

Affected Products

CXI SUPPORT CATHETER

Reason

This recall is being initiated by Cook Medical in response to six product complaints associated with reports of the 2.6 Fr curved catheter being mixed with the 4.0 Fr straight catheter and vice versa. Investigation revealed the products were mixed during processing.

Affected products

CXI SUPPORT CATHETER

Lot or serial number

5505259

Model or catalog number

CXI-2.6-18-90-P-NS-ANG

Companies
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES