CXI SUPPORT CATHETER (2015­-10-­22)

Starting date:
October 22, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55858

Affected Products

CXI SUPPORT CATHETER

Reason

This recall is being initiated by Cook Medical in response to six product complaints associated with reports of the 2.6 Fr curved catheter being mixed with the 4.0 Fr straight catheter and vice versa. Investigation revealed the products were mixed during processing.

Affected products

CXI SUPPORT CATHETER

Lot or serial number

5505259

Model or catalog number

CXI-2.6-18-90-P-NS-ANG

Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana