Custom Tubing Packs
Brand(s)
Summary
Contact the manufacturer if you need additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
CUSTOM PACK WITH SOFTLINE |
92285211 |
BO-HQV 59203 |
PERFUSION SET WITH SOFTLINE COATING |
92276990 |
BO-H 55815 |
PERFUSION TUBING PACK WITH BIOLINE COATING |
More than 10 numbers, contact manufacturer. |
BEQ-H 48715 |
PERFUSION SET WITH BIOLINE COATING |
More than 10 numbers, contact manufacturer. |
Not Available |
Issue
The tubing sets are designed and labeled for 2 years shelf-life from the manufacturing date. However, during a review of the technical documentation of tubing sets, it was found out that there is not sufficient evidence to demonstrate the 2 year shelf life of three 3rd party components integrated into the tubing sets.
Recall start date: 2021-11-05 |
What you should do
Contact the manufacturer if you need additional information.
Additional information
Details
MAQUET CARDIOPULMONARY GMBH KEHLER STRASSE 31, RASTATT GERMANY 76437 |
Get notified
Receive notifications for new and updated recalls and alerts by category.