Custom Pack with Softline, Quadrox-i Neonatal Oxygenator with Integrated Hardshell Reservoir (2019-05-12)
- Starting date:
- May 12, 2019
- Posting date:
- May 24, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70023
Last updated: 2019-05-24
Affected Products
- Custom Pack with Softline
- Quadrox-i Neonatal Oxygenator with Integrated Hardshell Reservoir
Reason
Maquet Cardiopulmonary has received complaints in relation to leakage at the QUADROX-i Neonatal blood inlet connector.
Internal investigations have revealed that the leakage only occurs if the enclosed reducing screw adapter 1/4"" x 3/16"" is screwed onto the blood inlet connector of the QUADROX-i Neonatal Oxygenator. The leakage occurs due to a non-conforming blood inlet connector that, in conjunction with the use of the screw reducing adapter 1/4"" x 3/16"", does not result in a tight connectoradapter connection. The combination of the reducing screw adapter 1/4"" x 3/16"" with the blood outlet connector or the use of the blood inlet connector directly to tubing does not exhibit any leakage.
Affected products
A. Custom Pack with Softline
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
BO-HQV 48412
Companies
- Manufacturer
-
MAQUET CARDIOPULMONARY GMBH
KEHLER STR. 31, RASTATT, GERMANY, 76437
RASTATT
76437
GERMANY
B. Quadrox-i Neonatal Oxygenator with Integrated Hardshell Reservoir
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
BO-VKMO 11000
VKMO 10000
VKMO 11000