Health product recall

Cusa Excel Straight 23khz Handpiece

Last updated

Summary

Product
Cusa Excel Straight 23khz Handpiece
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Cusa Excel Straight 23khz Handpiece

More than 10 numbers, contact manufacturer.

C2600

Issue

The decision to conduct a voluntary correction of the product was based on the following: through an internal investigation of customer complaints, it was identified that there is a potential for the Cusa Excel 23khz Handpiece housing to crack. In order to mitigate this issue the devices will be reworked through our service and repair process for the units manufactured prior to Nov 2023. Note: this defect does not exist in any handpiece produced and manufactured after November 2023.

Recall start date: June 11, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Integra Lifesciences (Ireland) Limited
Ida Business And Technology Park, Tullamore, Offaly, Ireland,
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75722

Get notified