Health product recall

CUSA Excel Plus System Nosecones (2013-12-06)

Starting date:
December 6, 2013
Posting date:
January 13, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37475

Recalled products

  1. CUSA Excel Plus System Nosecones 

Reason

The purpose of this field safety notice is to inform affected customers of the potential of the blue coag button on the CUSA Excel CEM Nosecone to stick in the 'activated' position and potentially cause burn.

Affected products

A. CUSA Excel Plus System Nosecones 

Lot or serial number

All lots

Model or catalog number
  • C6623
  • C6636
Companies
Manufacturer

Integra Lifesciences (Ireland) Ltd.

Ida Business and Technology Park

Sragh, Tullamore, Offaly

IRELAND