Health product recall

CUSA Clarity Console C7000 Operators Manual (2019-10-21)



Starting date:
October 21, 2019
Posting date:
April 15, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72709


Last updated:

Affected Products

CUSA Clarity Console C7000 Operators Manual

Reason

It was discovered through internal review that we need to revise the operator's manual to reflect the most accurate sterilization parameters and hence the dry time parameters have been changed from 30 minutes to 40 minutes for the CUSA Clarity 36khz handpiece and components consistent with the CUSA Clarity 23khz handpiece and components.

Affected products

CUSA Clarity Console C7000 Operators Manual

Lot or serial number

All lots.

Model or catalog number

C7000

Companies
Manufacturer

Integra Lifesciences (Ireland) Ltd.

Ida Business and Technology Park

Sragh, Tullamore, Offaly

IRELAND