This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CT/MR Fletcher Set (2014-03-17)

Starting date:
March 17, 2014
Posting date:
April 14, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39005

Recalled Products

CT/MR Fletcher Set

Reason

The CT/MR Fletcher set is designed for gynaecological brachytherapy (Radiation therapy) and mainly intented for the treatment of the cervix. During tests done at a customer site, it has been revealed by the customer's physicist that there is a slight deviation in the accuracy of the dwell position of the source.

Affected products

CT/MR Fletcher Set

Lot or serial number

Not applicable

Model or catalog number

SET0109

Companies
Manufacturer

Eckert & Ziegler Bebig GMBH

Robert-Rossle-Strabe 10

Berlin

13125

GERMANY