This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CT Express Injector Unit

Starting date:
March 27, 2017
Posting date:
April 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63010

Reason

A malfunction reported to Bracco Injeneering on March 17, 2017, was encountered during service activity tests without a patient. The service engineer successfully performed a simulated injection. Some moments later error code 2037 was displayed on CT Express temperature maintainer model and the injector unit motor started to run independently.

Affected products

CT Express Injector Unit

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

650195

Companies

Manufacturer
Bracco Injeneering S.A.
Avenue De Sevelin 46
Lausanne
1004
SWITZERLAND