Health product recall

Csk Electrode

Starting date:
July 10, 2017
Posting date:
August 15, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64228

Reason

There have been complaints that after multiple reprocessing cycles, the epoxy resin which holds the TCN electrode can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device. The failure to completely remove any blood and/or tissue from the TCN electrodes prior to re-sterilization creates a potential risk of infection and spread of blood borne pathogens to patients.

Affected products

Csk Electrode

Lot or serial number

Not Applicable

Model or catalog number

TCN-10

Companies

Manufacturer

COSMAN MEDICAL, INC.

22 Terry Avenue

Burlington

01803

Massachusetts

UNITED STATES