This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CRW Systems

Starting date:
August 13, 2012
Posting date:
September 3, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15551

Recalled Products

A. CRW System - Intubation Head Ring Kit
B. CRW System - Lightweight Arc

Reason

The recall is to inform affected customers of a possible interference between the T-Handle on the HRAIM and the CRW Arc System, and rework affected HRAIM devices by providing a new design HRIMTSC T-Handle screw.

Affected products

A. CRW System - Intubation Head Ring Kit

Lot or serial number

All Lots

Model or catalog number

HRAIM, HRIMTS

Companies
Manufacturer
Integra Burlington MA Inc.

B. CRW System - Lightweight Arc

Lot or serial number

All Lots

Model or catalog number

CRW-ASL

Companies
Manufacturer
Integra Burlington MA Inc.